• Nominal frequency:
5 GHz (option: 10 GHz*)
  • Measurement principle:
Mode splitting in cavity–sphere coupling; measure when split modes 1 & 2 are symmetric.
  • Magnetic bias:
Hand-held tunable magnet
  • Sample type:
Low-magnetic-loss ferrite spheres
  • Sample diameter:
D < 1.5 mm (quartz tube holder)
  • Instrumentation:
VNA (e.g., Keysight PNA-X N5245A) or Q-Meter
  • Temperature:
Room temperature
  • Software:
Electrodynamic-model-based extraction of ΔH from split-mode frequency and Q-factor
  • Measurability examples:
  • ΔH = 0.2 Oe, D = 0.5 mm, Ms = 1800 Gs → Yes
  • ΔH = 3 Oe, D = 1.4 mm, Ms = 1200 Gs → Yes
  • ΔH = 16 Oe, D = 1.2 mm, Ms = 1700 Gs → Yes
  • ΔH = 20 Oe, D = 1.0 mm, Ms = 1800 Gs → No (no mode splitting)
  • Full setup: PC with dedicated software + VNA or Q-Meter; strong coupling is required for mode splitting.

* Customizations are possible (e.g., a 10 GHz cavity). The measurement is performed when split modes are symmetric; broad-linewidth samples that do not split cannot be characterized.